info@labwan.co.in +91 9056603301 Mon–Sat: 9:00 AM – 6:00 PM
Spectrophotometer

LABWAN Nano Drop / Microvolume Spectrophotometer LNS-5100

Model LNS-5100 In Stock

A micro-volume UV spectrophotometer designed for the measurement of nucleic acids and proteins, with additional bacterial cell concentration detection at OD600. It features an Android-based operating system with a built-in touchscreen and supports direct data output through USB and an SD-RAM card.

Get Quote Catalog Not Available
ISO 9001 Certified Service & Support
Key Features
  • Micro-volume measurement of nucleic acids and proteins.
  • Bacterial cell concentration detection using OD600.
  • Android operating system with 7-inch touchscreen.
  • Built-in printer for data printing.
  • Data output through USB memory and SD-RAM card.
  • Three optical path detection technology.
  • LED light source.
  • Sample measurement without dilution.
  • Fluorescence function.
  • Strong repeatability and high linearity

Specifications

Parameter Value
Optical Path 0.05mm (high concentration); 0.2mm, 1.0mm (ordinary concentration)
Detectors Ultraviolet silicon
Nucleic Acid Detection Range 15000 ng/uL (dsDNA)
Fluorescence Sensitivity DsDNA: 0.5pg/ul
Fluorescence Linearity <1.5%
OD600 Absorbance Range 0~6.000Abs
OD600 Absorbance Stability [0,3)<0.5%, [3,4)<2%
OD600 Absorbance Accuracy (0,2)<0.005A, [2,3)<1%, [3,4)<2%
OD600 Repeatability of Absorbance [0,3)<0.5%, [3,4)<2%
Noise Level 0.002Arms (OA, 260nm); Between peaks 0.005Arms (260nm)
Absorbance Range 0.04-300A (Equivalent to 10mm)
Absorbance 0.003Abs (0.2mm optical range)
Absorbance Accuracy ±1% (7.332Abs at 260nm)
Sample Volume Requirements 0.5~2ul
Wavelength Range 260nm, 280nm
Wavelength Accuracy ±1nm within
Wavelength Reproducibility <0.2nm
Sample Base Material Quartz fiber and stainless steel
Testing Time <6S
Power Consumption at Standby 5W
Data Output Method USB
Light Source LED
Power Adapter 24V DC
Power Consumption 25W
Dimension W200 x D260 x H165mm
Weight 4.5Kg